Safety and efficacy of a metal stent covered with a silicone membrane containing integrated silver particles in preventing biofilm and sludge formation in endoscopic drainage of malignant biliary obstruction: a phase 2 pilot study

Metadata Downloads
Abstract
BACKGROUND AND AIMS: Membrane-covered self-expandable metal stents (SEMSs) have been developed to prolong the patency of stents by reducing tissue hyperplasia or tumor ingrowth. However, their effectiveness is attenuated by stent clogging as a result of biofilm formation on the inner surface of the membrane. The aim of this pilot study was to evaluate the efficacy and safety of SEMSs covered with a silicone membrane containing integrated silver particles (Ag-P) in malignant distal biliary obstruction. METHODS: Twenty-four patients who underwent SEMS placement because of malignant distal biliary obstruction were enrolled in this single-center pilot study. The main outcomes were technical success, clinical success, adverse events, stent patency, and survival. RESULTS: The technical and clinical success rates were 100% and 91.7% (22 of 24), respectively. The rates of early and late adverse events were 22.7% and 36.4%, respectively. The primary reintervention rate was 27.3% (6 of 22). Only 1 case involving stent malfunction was associated with sludge impaction. Median stent patency was 179 days. During follow-up, there were no serious adverse events or mortality related to the stents or Ag-P. Serum and urine silver concentrations before and after stent placement and at 32 weeks after placement did not differ. All serum and urine silver concentrations were<3 μg/L (3 ppb) and 5 μg/L (5 ppb), respectively. CONCLUSIONS: SEMSs covered with a silicone membrane containing integrated Ag-Ps may be effective and safe in malignant distal biliary obstruction. Stent dysfunction related to sludge impaction may be less frequent using this new stent. (Clinical Research Information Service identifier: KCT 0002310.).
All Author(s)
T. H. Lee ; M. K. Jung ; T. K. Kim ; C. G. Pack ; Y. K. Park ; S. O. Kim ; D. H. Park
Intsitutional Author(s)
이태훈
Issued Date
2019
Type
Article
Keyword
Aged; Ampulla of Vater; Bile; Biofilms; Carcinoma, Hepatocellular; Cholangitis; Cholestasis; Common Bile Duct Neoplasms; Drainage; Endoscopy, Digestive System; Female; Gallbladder Neoplasms; Humans; Liver Neoplasms; Male; Metal Nanoparticles; Microscopy, Electron, Scanning; Middle Aged; Palliative Care; Pancreatic Neoplasms; Pancreatitis; Pilot Projects; Postoperative Complications; Prosthesis Failure; Self Expandable Metallic Stents; Silicones; Silver
Publisher
American Society for Gastrointestinal Endoscopy
ISSN
0016-5107
Citation Title
Gastrointestinal Endoscopy
Citation Volume
90
Citation Number
4
Citation Start Page
663
Citation End Page
672, e2
Language(ISO)
eng
DOI
10.1016/j.gie.2019.06.007
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/2784
Appears in Collections:
소화기내과 > 1. Journal Papers
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.