SCHMC

Imaging-Guided Deployment of a Balloon-Based Vascular Closure Device in Routine Clinical Practice

Metadata Downloads
Abstract
This study evaluated the feasibility and safety of imaging-guided deployment of a Mynx Control device (Cordis, Miami Lakes, Florida) at the femoral access. A total of 201 Mynx devices were used under imaging guidance in 137 patients (mean age, 64.18 years ± 13.8; male, 68.6%) between June 2020 and February 2022. A 5-F sheath was used in 176 (87.6%) procedures, a 6-F sheath was used in 17 (8.5%) procedures, and a 7-F sheath was used in 8 (4%) procedures. Indicator errors occurred in 9 (4.5%) procedures. However, technical success was achieved via manipulation under imaging guidance in all procedures. The mean time to hemostasis was 3.3 minutes ± 0.5, and the clinical success rate was 100%. The mean length of hospital stay was 7 days ± 7.5. After 4 (2%) procedures, small hematomas reabsorbed spontaneously. Imaging-guided deployment of a Mynx Control device is a feasible and safe option for access site closure. Even after an indicator error, successful hemostasis was achieved via real-time interaction.
All Author(s)
W. H. Lee ; H. N. Lee ; S. S. Kim ; S. Lee ; Y. Cho ; S. J. Park
Intsitutional Author(s)
이형남; 김승수_영상의학과
Issued Date
2022
Type
Article
Publisher
Society of Cardiovascular and Interventional Radiology
ISSN
1051-0443 ; 1535-7732
Citation Title
Journal of vascular and interventional radiology
Citation Volume
33
Citation Number
11
Citation Start Page
1337
Citation End Page
1341
Language(ISO)
eng
DOI
10.1016/j.jvir.2022.07.024
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/3615
Appears in Collections:
영상의학과 > 1. Journal Papers
공개 및 라이선스
  • 공개 구분공개
  • 엠바고Forever
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.