SCHMC

Improving Accuracy and Reliability of Hearing Tests: Measurement Standards for Audiometric Devices

Metadata Downloads
Abstract
Pure-tone audiometry, using an audiometer, is the fundamental hearing test for diagnosing hearing loss. The requirements of the devices and the detailed process for calibrating the related equipment are described in international standards. However, traceable calibration and uncertainty evaluation processes are not widely accepted or applied to the qualification and maintenance of audiometric equipment. Here, we briefly review standard measurement systems for audiometric devices and introduce their calibration procedures. The uncertainty of each calibration process was investigated, and its impact on hearing test results was considered. Our findings show that the traceability of each procedure can be secured, satisfying the uncertainty requirement and being sufficiently smaller than the permissible deviation from the audiometer requirement. To guarantee the objectivity and reliability of hearing tests and maintain low uncertainty, close cooperation and mutual understanding between the metrology field and the medical community are necessary.
All Author(s)
Wan-Ho Cho ; Jihyun Lee ; Young Joon Seo ; Tae Hoon Kong ; Michelle J Suh ; In-Ki Jin ; Soo Hee Oh ; Hyo-Jeong Lee ; Seong Jun Choi ; Kyung-Ho Park ; Dongchul Cha
Intsitutional Author(s)
최성준
Issued Date
2024
Type
Article
Keyword
Audiometry; Calibration; Measurement standards; Traceability; Uncertainty
Publisher
대한청각학회
Korean Audiological Society
ISSN
2384-1621 ; 2384-1710
Citation Title
Journal of audiology & otology
Citation Volume
28
Citation Number
3
Citation Start Page
167
Citation End Page
175
Language(ISO)
eng
DOI
10.7874/jao.2024.00227
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/4678
Appears in Collections:
이비인후과 > 1. Journal Papers
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.